實驗室供電解決方案
實驗室供電系統
實(shi)驗建筑內部有各(ge)種(zhong)類(lei)型的(de)實(shi)驗室及儀(yi)器(qi)設備,供電系統除了(le)維持(chi)實(shi)驗室特定的(de)環境用電外,還要(yao)滿足(zu)現有及未來(lai)增加的(de)各(ge)種(zhong)儀(yi)器(qi)的(de)特殊(shu)用電要(yao)求,對于離心機、層析冷(leng)柜、低溫冰箱帶壓縮機之類(lei)的(de)儀(yi)器(qi),它們的(de)電機啟動(dong)所(suo)需要(yao)的(de)電流(liu)往(wang)往(wang)是工作(zuo)電流(liu)的(de)很多(duo)倍,在啟動(dong)瞬間往(wang)往(wang)會(hui)(hui)影響該線路的(de)電壓波(bo)動(dong),如果接在該線路上(shang)所(suo)用的(de)大功率儀(yi)器(qi)較多(duo),就會(hui)(hui)引起(qi)儀(yi)器(qi)工作(zuo)不(bu)正(zheng)常(chang);
微電(dian)(dian)(dian)(dian)(dian)子儀器如微生物電(dian)(dian)(dian)(dian)(dian)測(ce)試儀、分光光度計、計算機等(deng)對(dui)電(dian)(dian)(dian)(dian)(dian)源質量的(de)(de)(de)要求比較高,大(da)功(gong)率儀器的(de)(de)(de)頻頻啟動會產(chan)生脈(mo)沖電(dian)(dian)(dian)(dian)(dian)壓(ya),而這些脈(mo)沖電(dian)(dian)(dian)(dian)(dian)壓(ya)很容(rong)易損(sun)壞元件或引起讀數(shu)波動、數(shu)據丟失等(deng)故障。所(suo)以,對(dui)實(shi)(shi)驗(yan)建(jian)筑供電(dian)(dian)(dian)(dian)(dian)系(xi)(xi)統(tong)的(de)(de)(de)設計,與普通辦公(gong)室是不(bu)同(tong)(tong)的(de)(de)(de),實(shi)(shi)驗(yan)用電(dian)(dian)(dian)(dian)(dian)負荷,除了(le)必須(xu)預留足(zu)夠的(de)(de)(de)富余電(dian)(dian)(dian)(dian)(dian)量以滿(man)足(zu)未來發展的(de)(de)(de)需(xu)要,還必須(xu)提供不(bu)間(jian)斷的(de)(de)(de)穩壓(ya)電(dian)(dian)(dian)(dian)(dian)源。基于實(shi)(shi)驗(yan)室的(de)(de)(de)與眾(zhong)不(bu)同(tong)(tong),實(shi)(shi)驗(yan)建(jian)筑的(de)(de)(de)供電(dian)(dian)(dian)(dian)(dian)系(xi)(xi)統(tong)從電(dian)(dian)(dian)(dian)(dian)源、線路、照明(ming)、安全等(deng)方方面(mian)面(mian)都有其(qi)獨(du)特性。實(shi)(shi)驗(yan)建(jian)筑的(de)(de)(de)用電(dian)(dian)(dian)(dian)(dian)量通常(chang)是現有用電(dian)(dian)(dian)(dian)(dian)量的(de)(de)(de)2倍(bei)。
用電安(an)全的基本(ben)要(yao)素(su):電氣絕(jue)緣好,保(bao)證安(an)全距離,線路與插座容量與設備功率(lv)相適宜,不使用三無(wu)產品(無(wu)生產日(ri)期,無(wu)質量合(he)格證,無(wu)生產廠家(jia))
一、實驗室電源
不間斷的實驗室穩壓電源是保證儀器穩定運行的重要部分。為了避免市電的供電電壓不穩定或突然停電而影響實驗室的運行,通常加裝備用電源及穩壓器,最常用于實驗室的有UPS電源,分為兩類,一類是普通的UPS,只起到不間斷供電作用,但不能穩壓;另一類是在線式UPS電源,既能不間斷供電又能穩壓。可根據實際情況和儀器要求選擇不同的電源。建筑電源來源配電室,由甲方提供,電壓AC220/380V,50HZ,采用三相五線制配電系統,接地型式采用tn-s系統,要求中性線與保護線絕對分開并相互絕緣。要求使用選用阻燃性好的照明及電纜在施工前提前布置好,并嚴格使用。
二、實驗室供電線路:
1、為了使這些(xie)大(da)功率(lv)儀器(qi)工作時互不(bu)干(gan)擾,一般給大(da)功率(lv)儀器(qi)單獨設一條線(xian)路,微電(dian)子儀器(qi)與大(da)功率(lv)用電(dian)器(qi)不(bu)能共(gong)接(jie)同(tong)一條線(xian)路;
2、對于(yu)需(xu)要不(bu)間(jian)斷供電(dian)的精密儀器,應(ying)配(pei)穩壓(ya)的UPS電(dian)源;對于(yu)培養(yang)室(shi)(shi)、生物安全(quan)實驗(yan)室(shi)(shi)、動物飼養(yang)室(shi)(shi)等需(xu)要不(bu)間(jian)斷供電(dian)的實驗(yan)室(shi)(shi),必需(xu)采(cai)用(yong)雙保(bao)險專用(yong)電(dian)源。
3、每(mei)一(yi)實驗室(shi)內(nei)設(she)三相(xiang)交流電(dian)及單(dan)項交流電(dian),在靠近門口設(she)置設(she)一(yi)個總電(dian)源控制開關,方便從(cong)走廊引線(xian)、控制檢(jian)修(xiu)及開啟或切(qie)斷室(shi)內(nei)電(dian)源。對于(yu)實驗停止后仍須運行的(de)設(she)備,應連(lian)在專用供電(dian)電(dian)源的(de)線(xian)路上,避免因切(qie)斷實驗室(shi)的(de)總電(dian)源而(er)影響工作(zuo)。
4、實驗臺設(she)置一定數量的(de)三相及單相電(dian)(dian)源插座(zuo),電(dian)(dian)源插座(zuo)回路(lu)設(she)有漏(lou)電(dian)(dian)保(bao)護電(dian)(dian)器(qi),插座(zuo)設(she)置應遠(yuan)離水盆和(he)煤氣。
6、潮濕、有腐蝕性氣體、蒸(zheng)汽、火災危險和(he)爆炸危險等場所,應選用具(ju)有相應的防護性能的配電(dian)設備。
7、化學實驗(yan)室因有腐蝕性氣體,配電(dian)導線采用(yong)銅芯(xin)。物理實驗(yan)室可以采用(yong)鋁芯(xin)導線。
8、實驗室(shi)的接(jie)地(di)(di)系統可保(bao)證人身安全以(yi)及(ji)儀器的正常運轉。一般(ban)接(jie)地(di)(di)種(zhong)類有(you)安全保(bao)護(hu)接(jie)地(di)(di)、防靜電接(jie)地(di)(di)、直流接(jie)地(di)(di)、防雷(lei)接(jie)地(di)(di)等。
9、在同(tong)(tong)一(yi)(yi)(yi)科(ke)學實(shi)驗建筑(室(shi))內設(she)有兩(liang)種及以(yi)上不同(tong)(tong)電(dian)(dian)(dian)(dian)壓或頻(pin)(pin)率(lv)的(de)電(dian)(dian)(dian)(dian)源供電(dian)(dian)(dian)(dian)時(shi),宜分(fen)別(bie)設(she)置(zhi)配(pei)(pei)電(dian)(dian)(dian)(dian)保(bao)護(hu)裝置(zhi)并有明顯(xian)區分(fen)或標志。當(dang)由同(tong)(tong)一(yi)(yi)(yi)配(pei)(pei)電(dian)(dian)(dian)(dian)保(bao)護(hu)裝置(zhi)供電(dian)(dian)(dian)(dian)時(shi),應(ying)(ying)有良(liang)好的(de)隔離。不同(tong)(tong)電(dian)(dian)(dian)(dian)壓或頻(pin)(pin)率(lv)的(de)線(xian)路(lu)應(ying)(ying)分(fen)別(bie)單(dan)獨敷(fu)設(she),不得在同(tong)(tong)一(yi)(yi)(yi)管內敷(fu)設(she)。同(tong)(tong)一(yi)(yi)(yi)設(she)備(bei)或實(shi)驗流水線(xian)設(she)備(bei)的(de)電(dian)(dian)(dian)(dian)力線(xian)路(lu)和(he)無防干擾要求的(de)控制回路(lu)允許同(tong)(tong)一(yi)(yi)(yi)管內敷(fu)設(she)。
10、高層或線路較多的多層科學實驗建筑,垂直線路宜采用管道井敷設。強、弱電管線宜分別設置管道井。當在同一管道井內敷設時,應敷設在管道井內兩側。
三、實驗室照明
實驗室(shi)(shi)的照明設備一(yi)般以日(ri)光燈(deng)為宜,它不(bu)但(dan)使用壽命長,電(dian)源面積大,光效(xiao)高,而且發(fa)熱量低。在分析化(hua)學實驗室(shi)(shi),用目(mu)視法判斷容量滴定指示劑變(bian)色終點時,可(ke)在操作處安設熒光燈(deng);電(dian)磁干擾(rao)要求嚴(yan)格(ge)的實驗室(shi)(shi),不(bu)宜采(cai)用氣體放電(dian)燈(deng);
在暗室(shi)、電鏡室(shi)等應設單色(se)(se)(紅色(se)(se)或黃色(se)(se))照明,入口處宜(yi)設工作狀態標志燈;
放射性實(shi)驗室(shi)(shi)、傳(chuan)染性微生物(wu)實(shi)驗室(shi)(shi)以及從事致癌物(wu)或毒物(wu)操作的實(shi)驗室(shi)(shi),應采用嵌裝式潔凈燈(deng)具(ju),電線管路要力求(qiu)暗裝,電燈(deng)開關應裝在室(shi)(shi)外(wai)走(zou)廊(lang)上;
無菌(jun)室(shi)需要安(an)裝紫外(wai)滅菌(jun)燈,其控制開(kai)關(guan)應設(she)在門外(wai)并與一(yi)般照明(ming)燈具的控制開(kai)關(guan)分(fen)開(kai)設(she)置;
潮濕、有(you)腐蝕性氣體和(he)蒸汽、火災危(wei)險(xian)和(he)爆炸危(wei)險(xian)等場所,應(ying)選(xuan)用具有(you)相應(ying)防護性能的燈(deng)具;
在安全出口,疏散通道等處應設置安裝疏散指示燈,使疏散的人員能在緊急事故的情況下得以迅速疏散;管道技術層內應設照明并由單獨支路或專用配電箱(盤)供電。
四、實驗建筑照度標準表
五、實驗室用電的注意事項:
A.實驗(yan)前先(xian)檢查用電(dian)設備,再(zai)接通電(dian)源;實驗(yan)結束后,先(xian)關(guan)儀(yi)器(qi)設備,再(zai)關(guan)閉電(dian)源。
B.工(gong)作人員離(li)開實驗室(shi)或遇突然(ran)斷(duan)電,應關(guan)閉電源。
C.不得(de)將供電線(xian)任意放在通(tong)道上,以(yi)免因絕(jue)緣(yuan)破(po)損造成短路(lu)。
做完(wan)實(shi)驗或離開實(shi)驗室要(yao)及時(shi)斷電(dian),確保實(shi)驗裝置不帶(dai)電(dian)(例如儲能(neng)系統電(dian)壓應在安全電(dian)壓以下)。
更多實驗(yan)室整體規(gui)劃設計裝(zhuang)修請聯(lian)系科瓦特實驗(yan)室,0760-23626818.
QQ:1582339372
免費(fei)設計方案與報價:18022106989 熊(xiong)小姐
同類文章排行
- 實驗室規劃建設需要注意的地方
- 科瓦特2023年春節放假通知
- 成功簽約常熟立訊實驗室改造工程
- 喜訊,簽約完美生命科技研究院SPF動物實驗室
- 中山科瓦特國慶中秋放假通知
- 簽約中山板芙醫院PCR檢測室工程
- 喜訊 中標鴻基創能科技公司實驗室工程
- 成功簽約廣東威科質量檢測公司SPF級動物實驗室
- 2019年,科瓦特機電開工大吉
- 2019年春節放假通知
最新資訊文章
- 構建實驗室的前期了解和規劃
- 實驗室裝修規范與要求詳解
- 要建設一個實驗室,需要考慮的7點
- 實驗室裝修全攻略
- 實驗室裝修設計八大要點
- 建設一個實驗室,要涉及到哪些內容
- 實驗室裝修這些細節不能忽視!
- 實驗室設計關鍵的三要素,不可忽視!
- 實驗室建設設計的哪些細節是關鍵
- 實驗室建設設計的基本步驟有哪些?